Watch list

 
 
 
 
 
Differential pressure transmitter

P 26

High precision differential pressure transmitter with scalable measuring ranges

Also available as "air meter" to determine air conditioning consumption, see this chapter:
Measuring volume flow and air consumption



 

Differential pressure transmitter with scalable measuring range and a high level of overpressure protection

Special features

  • High level of accuracy and long-term stability through automatic zero-point calibration
  • Measurement of differential pressure or volumetric flow / flow rate possible
  • Easy configuration of the measuring range, analog output, time constants and units via PC or keyboard (optional)
  • USB interface (optional)
  • Display with keyboard for configuration of the pressure transmitter (optional)
  • Multilingual menu (English/German/Italian/French)
  • Suitable for top-hat rail mounting and wall surface installation
  • 2 floating switching relays (optional)
  • Integrated power adapter for 115/230 VAC power supply
  • Silicon-free materials for parts in contact with medium
  • Output signal can be configured 0 .. 10 V, 0/4 .. 20 mA or ± 5 V
  • Selectable attenuation of the output signal from 0 .. 60 s
  • Can be calibrated
  • Highly stable measurements thanks to cyclical self-calibration of the zero point
  • High level of overpressure protection even in the smallest measuring ranges

The P 26 differential pressure transmitter is designed for pressure monitoring applications in cleanrooms, laboratories, pharmaceutical plants and operating theatres as well as for special uses in machine and plant construction. As well as measuring the differential pressure, the instrument also records positive and negative overpressures as well as flow rates or volumetric flows.
The measuring range is scalable in the range from 10 .. 100 %. Two optional contact points allow the user to connect visual or audible alarms that are triggered any time values rise above or fall below specified limits. The transmitter can be configured easily via an USB interface (optional) with software. This can also be performed using a keyboard and display (optional). The measured values can be displayed in all standard units. A selectable time constant allows the transmitter to be adjusted perfectly to the pressure conditions and thus ensures a stable output signal. The P 26 can be mounted either on a wall or on top-hat rails.

Measurement ranges
10 / 50 / 100 / 250 / 500 Pa
1 / 2.5 / 5 / 10 / 20 / 50 / 100 kPa
freely scalable from 10 .. 100 % within a measurement range
(others available on request)

Margin of error
± 0,2 % or ± 0,5 % of the scaled range (40 .. 100 % of max. value, min. 0.3 Pa)

Output (linear / root-extracted)
0 .. 10 V (RL ≥ 2 kΩ)
0 .. 20 mA (RL≤ 500 Ω)
4 .. 20 mA (RL≤ 500 Ω)
± 5 V (RL ≥ 2 kΩ)
(output signals can be configured freely)

Power supply

24 VAC/DC, ± 10 %
24 VAC, +6 % (with galvanic seperation)
230/115 VAC, -15 %

Interface (optional)
USB, data cable supplied
External zero-point calibration

Display (optional)
multi-coloured LCD and keyboard

The instrument is available with an ISO factory calibration or DAkkS calibration certificate (optional).

 

Similar products

product picture

Differential pressure transmitter P 34

Find out more


product picture

Differential pressure transmitter P 29

Find out more


product picture

Differential pressure transmitter PU/PI

Find out more

 
 

© 2016 halstrup-walcher GmbH | Stegener Str. 10 | D-79199 Kirchzarten | Telefon +49 (0) 7661/3963-0 | info@halstrup-walcher.de | General terms and conditions | Imprint | Data protection